Problems with the Intoxilyzer 8000 in Florida

How Does the Intoxilyzer 8000 Work?

The Intoxilyzer 8000 uses infrared spectroscopy to analyze the alcohol content of a breath sample. Infrared spectroscopy uses duel wavelength detectors of infrared energy to analyze for ethanol. The machine introduces an infrared light source into the breath sample blown into the instrument. The hardware then samples that light by using an infrared detector to measure the samples. Measurements are taken many times each second.

The process of measuring the samples produces hundreds of pieces of "raw date." The "raw data" collected by the hardware is then interpreted by the software used by the Intoxilyzer 8000. The result of that software interpretation is then provided on a printout reading which estimates the breath alcohol level.

Four Requirements of Each Breath Test

The Intoxilyzer 8000 was developed with four requirements that are aimed at insuring the reading is scientifically relaible and accurately reflect the alcohol concentration. 

  1. The person submitting to the DUI breath test must blow with sufficient flow.
  2. The flow must be maintained continuously for a minimum time.
  3. A minimum volume of air must be blown into the intoxilyzer 8000;
  4. The slope detector, which monitors the rate of change in the alcohol concentration of the breath sample must be satisified.

Fail Safe Procedures used by the Intoxilyzer 8000

The software used by the Intoxilyzer 8000, in order to calculate a reliable reading, must utilizes several fail safe procedures. Without viewing the "source code" and software that runs the instrument, it can not be determined whether the fail safe procedures used by the Intoxilyzer 8000 result in an accurate reading. These fail safe procedures include the following:

  1. RFI Detect - whether a breath test result was affected by radio frequency;
  2. Interferent Detect Function - whether the reading detected by the infrared detector was caused by alcohol or by other substances in the breath sample;
  3. Slope Function - whether the alcohol found in the breath sample came from the defendant's blood or from mouth alcohol;
  4. Purge Failure - whether a breath result was contaminated by failing to purge the previous result from the sample chamber;
  5. Ambient Fail Error - whether a breath result was contaiminated by preexisting alcohol in the breath chamber;
  6. Control Tests - whether the machine was within calibration;
  7. In Range - whether each of the two BAL results were within 0.02 of each other;
  8. Volume Met - whether a sufficient volume of air was blown into the Intoxilyzer 8000 to obtain a reliable result (in Florida, the Intoxilyzer 8000 requires a breath sample of at least 1.1 liters).

Volume Not Met

The Florida version of the Intoxilyzer 8000 (which includes both the hardware and software that runs the machine), requires a breath sample of at least 1.1 liters. For any particular breath test sample, the machine should report to the breath test technican the warning "volume not met" when the breath sample is less than 1.1 liters. If the machine is operating correct, any breath test result that does not show a warning of "volume not met" should contain a sample with a volume greater than 1.1 liters. However, this fail safe procedure does not work in many instances.

One well documented flaw in the software can be demonstrated by the fact that the Intoxilyzer 8000 reported a volume of less than 1.1 liters but did not provide the "volume not met" warning. Without inspecting the "source code" and software it is impossible to determine the reliability of the reported volume measurement detected.

When questioned about this software flaw, officials for the Florida Department of Law Enforcement (FDLE) provided an explaination that sought to minimize the impact of the flaw. The FDLE reported that the flaw was limited to cases in which a person continued to give a breath sample for longer than three minutes into the test. By examining the test results, this interpretation of the software flaw is in incorrect because in some circumstances the machine would display the "volume not met" warning when the reported time for the breath sampe was four minutes.

Another well documented flaw in the software can be demonstrated by the fact that the Intoxilyzer 8000 reported a volume of more than 1.1 liters but still displayed the "volume not met" warning. If a problem exists with the reported volume of a breath sample, then the reliability of the corresponding breath alcohol level is also in question.  

 Perhaps the most problematic demonstration of a flaw in the software can be shown by the fact that the machine may report that no air was blown into the machine (0.0 liters) but still produce a breath alcohol level. Of course, if no sample breath is in the machine then no breath alcohol level could be detected. The only explanation for these facts is that a flow in the software is allowing the machine to registure a breath alcohol level where none should exist.

Software Flaws in the Intoxilyzer 8000

 

 

 

 

uses a computer program.

Sammis Law Firm, P.A.
1005 N. Marion St.  Tampa, FL 33602
Phone: (813) 250-0500 · Fax: (813) 314-9626
Map · Get Directions

The hiring of a DUI defense lawyer in Tampa, Florida, is an important decision that should not be based solely upon attorney ads, internet websites, informational videos, or blogs. Before you decide which Hillsborough County DUI Attorney to hire for your particular misdemeanor or felony case, ask us to send you additional free written information about our education, qualifications and experience. The law is constantly changing and developing, particular as it relates to drunk driving cases. Our internet website is for general information purposes only. The information presented in this website should not be construed to be formal legal advice, nor does it create any relationship between an attorney and a client. For legal advice, contact us today to request additional written information without cost or obligation or to speak directly with an attorney about your case.